User’s Guide32 Watt Temperature Control ModuleTCM-39032ILX Lightwave Corporation P. O. Box 6310 Bozeman, MT, U.S.A. 59771 :· · · 1-800-459-9459 ·
WARRANTYviii TCM-39032 in a container with at least 3 inches (7.5 cm) of compressible packaging material on all sides.We look forward to serving you
TCM-39032 1CHAPTER 1 INTRODUC
INTRODUCTION AND SPECIFICATIONSSpecifications12 TCM-39032CHAPTER 1 Specifications11. All values relate to a one-hour warm-up period.2. Output curre
INTRODUCTION AND SPECIFICATIONSSpecifications106_03 TCM-39032 3 CHAPTER 15. Temperature control range depends primarily on the type of thermistor a
INTRODUCTION AND SPECIFICATIONSInstallation4 TCM-39032CHAPTER 1 InstallationThis section describes the procedures for installing and removing a TCM
TCM-39032 5CHAPTER 2 OPERATIO
OPERATIONSENSOR SELECT Switches6 TCM-39032CHAPTER 2 SENSOR SELECT SwitchesThe SENSOR SELECT and THERM RANGE switches are used to select sensor type
OPERATIONTEC Connector06_03 TCM-39032 7 CHAPTER 2TEC ConnectorAt the right of center, on the TCM-39032 back panel, the user will find the 15-pin D-
OPERATIONBooster Operation8 TCM-39032CHAPTER 2 present, the TEC Interlock Error condition / event will be reported in the TEC Event Status REgister
OPERATIONBooster Operation06_03 TCM-39032 9 CHAPTER 2
OPERATIONBooster Operation10 TCM-39032CHAPTER 2
TCM-39032 9CHAPTER 3 CALIBRAT
CALIBRATIONEnvironmental Conditions10 TCM-39032CHAPTER 3 Environmental ConditionsCalibrate this instrument under laboratory conditions. ILX recomme
CALIBRATIONTEC Calibration Procedures06_03 TCM-39032 11 CHAPTER 35 Press and hold the (PARAMETER) SET switch and turn the ADJUST knob until the TEC
CALIBRATIONTEC Calibration Procedures12 TCM-39032CHAPTER 3 the GPIB hardware. Refer to the GPIB manual for details.7 Switch the THERM RANGE switch
CALIBRATIONTEC Calibration Procedures06_03 TCM-39032 13 CHAPTER 3If this value is to be measured and entered remotely via a GPIB controlled DMM, fo
CALIBRATIONTEC Calibration Procedures14 TCM-39032CHAPTER 3 Remote Operation LM335 Sensor The following procedure is for calibrating the LM335 senso
CALIBRATIONTEC Calibration Procedures06_03 TCM-39032 15 CHAPTER 3Local Operation RTD Sensor Calibration (with TSC-595 Option)The following procedur
CALIBRATIONTEC Calibration Procedures16 TCM-39032CHAPTER 3 Remote Operation RTD Sensor Calibration (with TSC-595 Option)The following procedure is
CALIBRATIONTEC Calibration Procedures06_03 TCM-39032 17 CHAPTER 3Local Operation ITE Current CalibrationThe following procedure is for calibrating
TABLE OF CONTENTS4_03 TCM-39032 i
CALIBRATIONTEC Calibration Procedures18 TCM-39032CHAPTER 3 Remote Operation ITE Current CalibrationThe following procedure is for calibrating the I
CALIBRATIONTEC Calibration Procedures06_03 TCM-39032 19 CHAPTER 3After the last “TEC:ITE” value is entered, the “*OPC?” query may be used to determ
CALIBRATIONTEC Calibration Procedures20 TCM-39032CHAPTER 3
TABLE OF CONTENTSii TCM-39032Remote Operation Thermistor Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . 11Local Operation AD590 S
TCM-39032 iiiSAFETY AND WARRANTY INFORMATIONThe Safety and Warranty Information section provides details about cautionary symbols used in the manual
SAFETY SYMBOLSiv TCM-39032 SAFETY SYMBOLSThis section describes the safety symbols and classifications. Technical specifications including electrica
WARRANTY06_03 TCM-39032 v WARRANTYILX LIGHTWAVE CORPORATION warrants this instrument to be free from defects in material and workmanship for a perio
WARRANTYvi TCM-39032 Claims for Shipping DamageWhen you receive the instrument, inspect it immediately for any damage or shortages on the packing li
WARRANTY06_03 TCM-39032 vii Comments, Suggestions, and ProblemsTo ensure that you get the most out of your ILX Lightwave product, we ask that you di